Date: |
2018-04-04 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Calibration Location: Lab
Last calibration: 2017-11-27
RSSI=73%, Optical RH=14%
Ambient P=100.7kPa, Tair=23.9C
Zero Air at 100psi, LBL Air 1.9497 at 1700
CH4 Zero Offset: -.002ppm
CH4 Span Offset: .167ppm
Notes: Powered up twice successfully and synced waveforms within 20 seconds. Actual laser temperature = 25.6C, set laser temperature = 26.C.
|
Entered by: |
|
|
|