Equipment and Calibration History

View Equipment |

View history entry 3351 for Licor LI-7500A sn: 75H-2150

Date: 2018-03-05
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-02-20 AGC=97 SDM=0 DAC1=0 DAC2=0 Ambient P=101.2kPa, Tair=25.7C, Tdew=1.7C Zero Air at 100psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.902 -> .9023 Offset: 1.32ppm CO2 span 1.007 -> 1.0093 Offset: -1.21ppm H2O zero 0.866 -> .9655 Offset: -3.3ppm H2O span 1.009 -> 1.0096 Offset: -.01ppm Notes:
Entered by: