Equipment and Calibration History

View Equipment |

View history entry 3347 for Licor LI-7500A sn: 0042

Date: 2018-03-05
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-05-05 AGC=98 SDM=0 DAC1=0 DAC2=0 Ambient P=101.4kPa, Tair=25.5C, Tdew=2.0C Zero Air at 150psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9847 -> .9485 Offset: 1.08ppm CO2 span 1.0244 -> 1.0124 Offset: -1.35ppm H2O zero 0.8434 -> .843 Offset: .7ppm H2O span 1.1404 -> 1.0225 Offset: -.11ppm Notes:
Entered by: