Equipment and Calibration History

View Equipment |

View history entry 3345 for Licor LI-7500A sn: 0041

Date: 2018-03-05
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-02-20 AGC=99 SDM=0 DAC1=0 DAC2=0 Ambient P=101.5kPa, Tair=23.4C, Tdew=10.6C Zero Air at 150psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9971 -> .9972 Offset: .25ppm CO2 span 1.0195 -> 1.0196 Offset: -1.07ppm H2O zero 0.8187 -> .8194 Offset: 4.5ppm H2O span 1.0217 -> 1.0267 Offset: -.1ppm Notes: H2O span checked against our dewpoint generator and Patty's dewpoint generator
Entered by: