Equipment and Calibration History

View Equipment |

View history entry 3335 for Licor LI-7500A sn: 0041

Date: 2018-02-20
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-09-11 AGC=98.4 SDM=0 DAC1=0 DAC2=0 Ambient P=101.1kPa, Tair=23.8C, Tdew=-2.4C Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9969 -> .9971 Offset: .92ppm CO2 span 1.0155 -> 1.0195 Offset: -1.93ppm H2O zero 0.82 -> .8187 Offset: -7.5ppm H2O span 1.0217 -> 1.0217 Offset: 0ppm Notes: I didn't do a H2O span because the sensor's dewpoint measurement was hovering around 2C, and then dropping towards 0C over 30 minutes. There may be a problem with the dewpoint generator.
Entered by: