View history entry 3333 for Licor LI-7500A sn: 75H-2150

Date: 2018-02-20
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-10-11 AGC=96.9 SDM=0 DAC1=0 DAC2=0 Ambient P=101.2kPa, Tair=24.4C, Tdew=0.15C Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9018 -> .902 Offset: .58ppm CO2 span 1.007 -> 1.007 Offset: .83ppm H2O zero 0.8649 -> .866 Offset: 7.1ppm H2O span 1.009 -> 1.009 Offset: 0ppm Notes: I didn't span the H2O because the sensor's dewpoint measurement decreased from 10 to 4.3 after 50 minutes. It didn't make sense for the H2O span to change that much. There might be an issue with the dewpoint generator.