Equipment and Calibration History

View Equipment |

View history entry 3280 for Licor LI-7500A sn: 75H-2180

Date: 2018-01-10
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-08-16 AGC=97.8 SDM=0 DAC1=0 DAC2=0 Ambient P=101.1kPa, Tair=17.4C, Tdew=10.6C Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9129 -> 0.9147 Offset: 7.92ppm CO2 span 1.0145 -> 1.0362 Offset: -10.74ppm H2O zero 0.8871 -> 0.8864 Offset: -4.2ppm H2O span 1.0164 -> 1.0168 Offset: -0.02ppm Notes: