Date: |
2018-01-10 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Adjusted |
Notes: |
Checked the sensor in the lab, and the temperature reading was fine. However, there was an error "Reference signal is UNLOCKED," and CH4 was reading ~0.05 umol/mol. The diagnostics tab showed that the signal and reference waveforms were not synced. I changed the temperature control to manual, adjusted the laser temperature from 30.6 to 26.0 so that the waveforms were synced, and changed the temperature control back to automatic. |
Entered by: |
Daphne Szutu |
|
|