Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: cal_KippZonen_PARLite_sn030380_20031217.pdfReport: PAR comparison 12-2018 - 2018-12-11
Report: Oxford Tract Radiometer Comparison - 06/2019 - 2019-06-18
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
4432 | 2020-04-17 | Lab none | Installed | Bouldin Alfalfa PAR Sensor Under-A | Installed at BA at ground surface protected by a PVC collar to try to measure below canopy PAR. |
4072 | 2019-08-16 | Lab none | Calibration | Lab none | New calibration factor of 5.26 uV/(umol/m2/s). Tested this sensor at Oxford Tract against AMP reference sensor K&Z PQS1 sn 161073. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
3774 | 2018-10-30 | Lab none | Location | Lab none | This sensor has been on the tram in the incoming position for at least the last 10 years. |
4073 | 2003-12-17 | Factory | Calibration | Lab none | Initial factory calibration 5.31 uV/(umol/m2/s) |