Equipment and Calibration History

View Equipment |

View history entry 3201 for Licor LI-7500A sn: 0073

Date: 2017-11-27
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-06-22 AGC=100.6 SDM=0 DAC1=0 DAC2=0 Ambient P=101.0kPa, Tair=21.6C, Tdew=6.8C Zero Air at 350psi, LBL Air 408.97 at 1750 Tdewspan=16.00 CO2 zero 0.9572 -> .9571 Offset: -0.028ppm CO2 span 1.0084 -> 1.0118 Offset: -2.11ppm H2O zero 0.7307 -> .7306 Offset: -0.3ppm H2O span 1.0045 -> 1.0000 Offset: 0.1ppm Notes: Calibrated on 7550 software version 8.5.0
Entered by: