Equipment and Calibration History

View Equipment |

View history entry 3192 for Licor LI-7500A sn: 75H-2176

Date: 2017-11-09
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-06-12 AGC=89.6 SDM=0 DAC1=0 DAC2=0 Ambient P=100.6kPa, Tair=22.5C, Tdew=12.2C Zero Air at 400psi, LBL Air 408.97 at 1800 Tdewspan=16.00 CO2 zero 0.869 -> 0.869 Offset: 0ppm CO2 span 1.016 -> 1.0225 Offset: -3.37ppm H2O zero 0.8572 -> .8570 Offset: -1.3ppm H2O span 1.0022 -> 1.0047 Offset: -0.06ppm Notes: Calibrated on AIU-0674 (version 8.5.0)
Entered by: