Equipment and Calibration History

View Equipment |

View history entry 3190 for Licor LI-7700 sn: TG1-0357

Date: 2017-11-09
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2017-03-23 RSSI=72.7%, Optical RH=23% Ambient P=100.6kPa, Tair=21.8C Zero Air at 400psi, LBL Air 1.9497 at 1800 CH4 Zero Offset: -0.1ppm CH4 Span Offset: -0.0197ppm Notes: Optics RH was 23%, so I flushed the head cavity with zero air and replaced the desiccant. We just changed the desiccant last time (March 2017) so I was expecting a RH ~5%. Let's keep an eye on the optics RH.
Entered by: