Equipment and Calibration History
View Equipment |
Date: |
2010-05-05 |
From: |
Lab - none |
To: |
Twitchell Rice - Gas Analyzer CH4 |
Title: |
Installed |
Notes: |
Received from Liukang Xu of licor. Installed on the Twitchell tower. Single ended analog outs are being used to record the LI7700 data on the CR1000:
Channel 0V 5V wire
CH4 (mmol/m3) 0.05 0.15 wht
CH4 (ppm) 1 5 brn
Pressure (kPa) 90 110 grn
Temperature (C) 0 50 yel
RSSI (%) 0 100 gry
The mirrors are set to clean at an RSSI value of 20%.
USB logging is set up at 10Hz. |
Entered by: |
|
|
|
Field notes