Equipment and Calibration History

View Equipment |

View history entry 3052 for Licor LI-7500A sn: 0041

Date: 2017-09-11
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-05-05 AGC=98.4 SDM=0 DAC1=0 DAC2=0 Ambient P=100.2kPa, Tair=26.5C, Tdew=13.0C Zero Air at 900psi, AMP Air 408.97 at 2075 Tdewspan=16.00 CO2 zero 0.9969 -> .9969 Offset: -.01ppm CO2 span 1.0177 -> 1.0155 Offset: 1.05ppm H2O zero 0.821 -> .82 Offset: -5.4ppm H2O span 1.032 -> 1.0217 Offset: .22ppm Notes:
Entered by: