Date: |
2010-06-16 |
From: |
Twitchell Rice - Gas Analyzer CH4 |
To: |
Twitchell Rice - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Clibrated in the field with grade 5.0 ultra pure N2 and 3.276ppm CH4 secondary standard.
zero before: ~0.009 ppm
zero after: ~0.000 ppm
span before: ~3.333 ppm
span after: ~3.277 ppm
|
Entered by: |
|
|
|