Equipment and Calibration History
View Equipment |
Date: |
2009-12-21 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Tested |
Notes: |
Tested on a CR1000 against an HMP45AC (sn U3030041) freshly returned from Vaisala factory calibration. See file HMP45A_CalCheck.xls. The comparison was for a very small range of temperatures ove a short time but;
Air Temperature: y = 0.7425x + 4.8166 R2 = 0.9972
Relative Humidity: y = 0.99x - 1.4165 R2 = 0.9986 |