Equipment and Calibration History

View Equipment |

View history entry 2941 for Licor LI-7500A sn: 75H-2667

Date: 2017-08-01
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-03-21 AGC=100 SDM= DAC1= DAC2= Ambient P=100.1kPa, Tair=23.2C, Tdew=19.4C Zero Air at 1200psi, LBL Air 408.97 at 2100 Tdewspan=16.00 CO2 zero 0.9423 -> .9443 Offset: 8.88ppm CO2 span 1.0144 -> .9346 Offset: 59.3ppm H2O zero 0.838 -> .8578 Offset: 144.2ppm H2O span 1.0219 -> .9724 Offset: 1.09ppm Notes: Reset signal strength from 113 to 100. H2O at ambient was about 900 mmol/m3, which seems higher than usual. Manually entered in existing coefficients before the calibration because did not have existing config file (usual laptop was elsewhere). Maybe I had entered the coefficients incorrectly in the database?
Entered by: