View history entry 294 for Licor LI-7500A sn: 0042

Date: 2011-02-02
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration Nov 16, 2010 AGC=67, SDM=7, DAC1 0-1000 H2O, DAC2 10-25 CO2 P=101.84 kPa Tair=21.06C, Tdew=~-1C Zero air no scrub 1600psi, CO2 597.4ppm 150psi, Tdew=-7.1C Precal CO2 zero=0.9622, span=0.9827 Postcal CO2 zero=0.9605, span=0.9855 Precal H2O zero=0.7829, span=0.9693 Postcal H2O zero=0.7877, span=1.1467 CO2 zero off by -6.13ppm, span off by -2.3ppm H2O zero off by +30.7mmol/m3, span off by -2.4C The weather is warm and dry making the dew point very low and difficult to calibrate.