Equipment and Calibration History

View Equipment |

View history entry 2843 for Licor LI-7500A sn: 0073

Date: 2017-06-21
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-06-13 AGC=101.1 SDM= DAC1= DAC2= Ambient P=99.3kPa, Tair=24.2C, Tdew=12.7C Zero Air at 1450psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.9567 -> .9574 Offset: 2.42ppm CO2 span 1.0065 -> 1.0075 Offset: -.67ppm H2O zero 0.7342 -> .7305 Offset: -21.8ppm H2O span 1.0251 -> 1.0071 Offset: .4ppm Notes: Checking the zero and spans after upgrade from 7500 to 7500A. There are jumps in CO2 zero and span so I will check again tomorrow.
Entered by: