Equipment and Calibration History

View Equipment |

View history entry 2840 for Licor LI-7500A sn: 75H-2514

Date: 2017-06-20
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-03-14 AGC=102.2 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=25.7C, Tdew=13.2C Zero Air at 1500psi, NOAA Ait 402.69 at 1650 Tdewspan=16.00 CO2 zero 0.9046 -> .9041 Offset: -2.08ppm CO2 span 1.0139 -> 1.0115 Offset: 0.96ppm H2O zero 0.8368 -> .8378 Offset: 6.4ppm H2O span 1.0106 -> 1.0019 Offset: 0.20ppm Notes:
Entered by: