View history entry 284 for Licor LI-7700 sn: TG1-0155

Date: 2011-01-31
From: Lab - none
To: Lab - none
Title: Calibration
Notes: On 1/27/11, Dennis and Jaclyn tested the LI-7700 output against the calibration gases in the lab. 1. The tank with 4.982 ppm methane read 5.021 ppm on the LI-7700. 2. The zero air tank read 0.05 ppm on the LI-7700 (although low confidence that the gas is in fact zero methane. 3. The 1.8978 god bottle read 1.898 on the LI-7700. We chose not to calibrate the instrument since the readings were so close to the actual values.
Entered by: