Equipment and Calibration History

View Equipment |

View history entry 282 for Campbell Sci. CS547A sn: 5688

Date: 2011-01-27
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Jaclyn Hatala did a temperature correction calibration for the EC probe on 1/27/11. 1. First, I submerged the probe in water of 25C for 10 minutes and recorded the uncorrected EC values from the datalogger program MBCR23X_TOWERMET_V100. 2. Next, I submerged the probe in water of 20C for 10 minutes and recorded the uncorrected EC values. 3. I used the formula TC=100 * (C-C_25)/((t-25)*C_25) provided by the Campbell manual to calculate the temperature correction in %/C. The value I got was 0.786%/C. 4. I changed the MBCR23x_TOWERMET_V100 program to use the uncorrected EC to provide a corrected EC value (Cond_corr).
Entered by: