View history entry 280 for Licor LI-7700 sn: TG1-0155

Date: 2011-01-25
From: Factory -
To: Lab - none
Title: Returned
Notes: Returned from repair at Licor. Service note say: "The spars were knocked out of alignment. The spinning mirror bearings were knocked out of alignment. Replaced the motor and realigned everything. Tested in environmental chamber from -25C to 50C. Calibrated. Changes to settings: Zero = 23.0176 Span = 0.00010261 sampledcoffset = 430 samplescoffset = -965