Equipment and Calibration History
View Equipment |
Date: |
2011-01-25 |
From: |
Factory - |
To: |
Lab - none |
Title: |
Returned |
Notes: |
Returned from repair at Licor. Service note say: "The spars were knocked out of alignment. The spinning mirror bearings were knocked out of alignment. Replaced the motor and realigned everything. Tested in environmental chamber from -25C to 50C. Calibrated.
Changes to settings:
Zero = 23.0176
Span = 0.00010261
sampledcoffset = 430
samplescoffset = -965 |
Entered by: |
|
|
|