View history entry 2796 for Licor LI-7500A sn: 75H-2150

Date: 2017-05-30
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-03-14 AGC=97.2 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=19.9C, Tdew=12.2C Zero Air at 1050psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.9014 -> .9024 Offset: 4.23ppm CO2 span 1 -> 1.0194 Offset: -9.8ppm H2O zero 0.8651 -> .8664 Offset: 9.3ppm H2O span 1.0094 -> .9975 Offset: .26ppm Notes: