Date: |
2017-04-03 |
From: |
Lab - none |
To: |
Factory - |
Title: |
Serviced |
Notes: |
Shipped to Campbell Sci for calibration. CSI Technician notes:
Tested & inspected. 2x4 connectors have corrosion, need replaced. Needs new o-rings. While testing sensor head, also found temp chip out of spec, drifting high in high humidity. Temperature chips are no longer available for this sensor. Sensor can not be repaired. Dispose of at CSI. |
Entered by: |
Daphne Szutu |
|
|