Equipment and Calibration History

View Equipment |

View history entry 2743 for Licor LI-7500A sn: 0042

Date: 2017-05-05
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-08-15 AGC=99.9 SDM= DAC1= DAC2= Ambient P=100.2kPa, Tair=25.5C, Tdew=7.8C Zero Air at 1600psi, LBL Air 513.22 at 350 Tdewspan=16.00 CO2 zero 0.9504 -> .9847 Offset: -6.7ppm CO2 span 1.001 -> 1.0244 Offset: -15.52ppm H2O zero 0.8519 -> .8434 Offset: -54.8ppm H2O span 1.003 -> 1.1404 Offset: -2.72ppm Notes: First calibration since upgrade from LI-7500 to LI-7500A
Entered by: