Equipment and Calibration History

View Equipment |

View history entry 2738 for Gill WM 1590 sn: 170524

Date: 2017-05-02
From: Lab - none
To: Lab - none
Title: Initial settings
Notes: All Checksum fields report PASS. Anemometer Firmware Version: 2329-701 Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Output Message Format : M2 Polar Output Message Format : M1 UV(W) Resolution settings : Normal resolution Resolution settings : High resolution Output rate : P1 1 Hz Output rate : P8 20 Hz SOS and Sonic Temperature display : Neither SOS and Sonic Temperature display : SOS Analogue output mode : 0v to 5V Analogue output mode : -5V to 5V Output Channel 1 : Polar angle (wrap mode = 360 degree wrap) Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : Polar Magnitude (full range = 5m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 4 : Status (status mode = Good signal high, bad signal low) Output Channel 4 : Speed of Sound THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER ======================== Serial Number : W170524 ---- SW VERSION -------------------------------------------- Anemometer Firmware Version: 2329-701 ---- HW DETAILS OF ANEMOMETER ------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER ---------------------- Baud Rate : 19200 Output Message Format : M2 Polar Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : Normal resolution Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : Neither SOS and Sonic Temperature display : SOS Analogue output mode : 0v to 5V Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : Polar angle (wrap mode = 360 degree wrap) Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : Polar Magnitude (full range = 5m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Status (status mode = Good signal high, bad signal low) Output Channel 4 : Speed of Sound Synchronised Polling : Off
Entered by: Daphne Szutu