Date: |
2017-03-23 |
From: |
West Pond - Gas Analyzer CH4 |
To: |
West Pond - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: West Pond
Last calibration: 2016-10-11
RSSI=86.2%, Optical RH=16%
Ambient P=102.0kPa, Tair=17.6C
Zero Air at 750psi, Ameriflux CH4 4.0304 at 1100
CH4 Zero Offset: -0.041ppm
CH4 Span Offset: -0.0251ppm
Notes: Changed dessicant and optics RH = 27%. Flushed head with span gas and optics RH dropped to 16%.
|