Date: |
2017-03-23 |
From: |
Mayberry - Gas Analyzer CH4 |
To: |
Mayberry - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: Mayberry
Last calibration: 2016-10-11
RSSI=77.3%, Optical RH=17%
Ambient P=102.1kPa, Tair=17.0C
Zero Air at 750psi, Ameriflux CH4 4.0304 at 600
CH4 Zero Offset: 0.032ppm
CH4 Span Offset: -0.0277ppm
Notes: Changed dessicant and flushed out head with dry air.
|
Entered by: |
|
|
|