Date: |
2017-03-14 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration: 2017-01-30
AGC=99.6 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=28.2C, Tdew=12.8C
Zero Air at 1700psi, God Air 421.22 at 400 Tdewspan=16.00
CO2 zero 0.9392 -> 0.9392 Offset: 0.11ppm
CO2 span 1.012 -> 1.0131 Offset: 0.57ppm
H2O zero 0.9039 -> 0.9041 Offset: 1.0ppm
H2O span 1.0054 -> 1.0063 Offset: .01ppm
Notes: This sensor has been upgraded to firmware version 8.5.0. I connected it to Licor box sn AIU-0674 and used software version 8.5.0 to calibrate this sensor in order to calibrate the temperature coefficients (these coefficients are not calibrated in software version 8.0.0).
|
Entered by: |
|
|
|