View history entry 2643 for Licor LI-7500A sn: 75H-2176

Date: 2017-03-07
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-10-25 AGC=91.5 SDM=-- DAC1=-- DAC2=-- Ambient P=101.1kPa, Tair=21.5C, Tdew=--C Zero Air at 1700psi, God Air 421.22 at 375 Tdewspan=16.01 CO2 zero 0.8689 -> 0.8691 Offset: 0.83ppm CO2 span 1.0176 -> 1.0195 Offset: -1.04ppm H2O zero 0.8563 -> 0.8561 Offset: -0.7ppm H2O span 1.0134 -> 1.0102 Offset: -0.04ppm Notes: