View history entry 2640 for Forerunner-Eosense GP-001 sn: GP20140021

Date: 2017-02-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: y = 13.36*x The probe was tested against a new probe with decreasing CO2 from 2.4% to zero. This probe read significantly lower than the standard probe and showed some lag at higher change rates.
Entered by: Joe Verfaillie