Equipment and Calibration History

View Equipment |

View history entry 2621 for Gill WM 1590 sn: 074417

Date: 2017-02-17
From: Lab - none
To: Lab - none
Title: Upgrade
Notes: Upgraded firmware to correct for W bug. The upgrade failed the first time because the Power-on message had been turned off. Without the Power-on message, the upgrade wizard cannot check if the checksums are correct. After I turned on the Power-on message, I used the second upgrade (X074417-s.gcu) file to complete the upgrade. Old SN: Y074417 New SN: X074417 Upgrade file: Y074417-Y124215-m.gcu, X074417-s.gcu Old firmware: 2329-114-01 New firmware: 2329-116-01 All Checksum fields report PASS. Log file: WMUpgradeLog_X074417_170217_121544.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Output rate : P2 2 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ====DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X074417 ----SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-116 ----HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 ADC resolution of attached unit : 14 DAC resolution of attached unit : 14 Number of ADC channels : 4 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ----USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P2 2 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Fixed Field SOS and Sonic Temperature display : SOS Analogue input display : OFF Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound
Entered by: Daphne Szutu