View history entry 2587 for Licor LI-7500A sn: 0073

Date: 2017-01-18
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-12-05 AGC=58 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.46kPa, Tair=19.9C, Tdew=5.83C Zero Air at 1750psi, God CO2 421.22 at 400 Tdewspan=16.00 CO2 zero 1.0707 -> 0.9573 Offset: -519ppm CO2 span 1.0029 -> 1.0037 Offset: -0.22ppm H2O zero 0.7391 -> 0.7394 Offset: -1.3ppm H2O span 1.0255 -> 1.0063 Offset: 0.41ppm Notes: Something happened to the CO2 coefficients after the last calibration causing the CO2 channel to be way off. New calibration seems good and matches historic coefficients.