View history entry 2529 for Licor LI-7500A sn: 75H-2668

Date: 2016-12-06
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-05-20 AGC=103.3 SDM=6 DAC1= DAC2= Ambient P=100.29kPa, Tair=15.36C, Tdew=3.00C Zero Air at 350psi, God CO2 421.22 at 400 Tdewspan=16.02 CO2 zero 0.9385 -> 0.9385 Offset: 0.0ppm CO2 span 1.0136 -> 1.0171 Offset: -1.92ppm H2O zero 0.8969 -> 0.9017 Offset: 30.2ppm H2O span 1.0146 -> 1.0841 Offset: -1.41ppm Notes: