Date: |
2016-10-27 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Calibration Location: Lab
Last calibration: 2014-03-07
RSSI=79.3%, Optical RH=17%
Ambient P=100.4kPa, Tair=22.8C
Zero Air at 400psi, CH4 2nd 4.995 at 500
CH4 Zero Offset: 0.042ppm
CH4 Span Offset: 0.057ppm
Notes: Last calibration at Licor after replacement of laser due to unknown failure.
|
Entered by: |
|
|
|