Equipment and Calibration History

View Equipment |

View history entry 2516 for Licor LI-7500A sn: 75H-2176

Date: 2016-10-25
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-06-03 AGC=91.2 SDM=6 DAC1=- DAC2=- Ambient P=100.28kPa, Tair=20.11C, Tdew=12.3C Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.01 CO2 zero 0.8688 -> 0.8689 Offset: 0.22ppm CO2 span 1.0175 -> 1.0176 Offset: -0.12ppm H2O zero 0.8571 -> 0.8563 Offset: -4.9ppm H2O span 1.0281 -> 1.0134 Offset: 0.31ppm Notes: