Date: |
2016-10-25 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Maintenance |
Notes: |
This sensor was serviced in the lab. The probe was taken apart and cleaned with vinegar and DI water. The membrane was replace and the probe was reassembled with fresh solution. In ambient air it was reading 80.0mV |
Entered by: |
Joe Verfaillie |
|
|