View history entry 2396 for Licor LI-7500A sn: 0042

Date: 2016-08-15
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-02-22 AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.52kPa, Tair=22.05C, Tdew=12.36C Zero Air at 500psi, God CO2 421.22 at 500 Tdewspan=16.00 CO2 zero 0.9505 -> 0.9504 Offset: -0.42ppm CO2 span 1.0107 -> 1.0010 Offset: 0.63ppm H2O zero 0.8522 -> 0.8519 Offset: -1.8ppm H2O span 0.999 -> 1.0030 Offset: -0.08ppm Notes: Exterior serial connector did not work.
Entered by: