Date: |
2016-08-11 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Calibration Location: Lab
Last calibration: 2015-08-10
RSSI=80.6%, Optical RH=27%
Ambient P=100.6kPa, Tair=22.2C
zero air at 500psi, AMP CH4 1.9966 at 400
CH4 Zero Offset: .029ppm
CH4 Span Offset: 0.0006ppm
Notes: First lab calibration since receiving this instrument. The desiccant was replaced - instrument was sitting idle in lab for almost a year.
|
Entered by: |
|
|
|