Date: |
2016-07-07 |
From: |
Tonzi - |
To: |
Tonzi - |
Title: |
Calibration |
Notes: |
Last calibration was 2016-06-16 Cell T = 52.2, Pressure = 91.0
Zero Air offset = -2.0ppm
Span CO2 405ppm offset = 5.0ppm
Zero 0.9472 -> 0.9467
Span 0.8496 -> 0.8369 |
Entered by: |
Joe Verfaillie |
|
|