Date: |
2016-05-26 |
From: |
Tonzi-Tower - |
To: |
Tonzi-Tower - |
Title: |
Calibration |
Notes: |
Last calibration was 2016-02-09
Cell T = 52.2, Pressure = 92.0
Zero Air offset = 16.4ppm
Span CO2 405ppm offset = -19.4ppm
Zero 0.9423 -> 0.9461
Span 0.8220 -> 0.8499
Changed Pressure output range from 1.33331998E3 to 115kPa |
Entered by: |
Joe Verfaillie |
|
|