Equipment and Calibration

View History

Equipment Entry

Instrument:Flowmeter
Manufacture:Cole Parmer
Model No:32707-30
Serial No:CP6295
Tag No:
Acquisition Date:2011-11-01
Project:SPUR
Location on 2011-11-04:Lab,
Notes: 0-1000ml/min

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
446 2011-10-20 Factory Calibration Factory Initial calibration. Flow rates at STP, max allowable error +-1.5%, Cal for Air, at 23C, 740mmHg, 12.8VDC

ml/min = 202.56 * VDC - 8.8158, R2 = 0.9998

ml/min = 1.0115 * screen - 7.7203, R2 = 0.9998

Cal sheet on file.