Equipment and Calibration History

View Equipment |

View history entry 2259 for Licor LI-7500A sn: 0041

Date: 2016-05-05
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2015-10-28 AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.03kPa, Tair=19.99C, Tdew=9.10C Zero Air at 600psi, God CO2 421.22 at 500 Tdewspan=11.09 CO2 zero 1.0004 -> 1.0002 Offset: -0.7ppm CO2 span 1.0107 -> 1.0128 Offset: -1.12ppm H2O zero 0.8247 -> 0.8257 Offset: 5.0ppm H2O span 0.8836 -> 1.0037 Offset: -2.5ppm Notes: Large offset in H2O span. I think a mistake was made last time when the dew point was set to the H2O mmol/mol value.
Entered by: