Equipment and Calibration History

View Equipment |

View history entry 2258 for Licor LI-7500A sn: 0073

Date: 2016-05-05
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2015-10-28 AGC=58 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=99.90kPa, Tair=19.2C, Tdew=8.98C Zero Air at 600psi, God CO2 421.22 at 500 Tdewspan=11.09 CO2 zero 0.9584 -> 0.9579 Offset: -1.8ppm CO2 span 1.0034 -> 1.0029 Offset: 0.18ppm H2O zero 0.7403 -> 0.7385 Offset: -9.8ppm H2O span 0.9201 -> 1.0159 Offset: -2.01ppm Notes: Large offset in H2O span. I think a mistake was made last time when the dew point was set to the H2O mmol/mol value.
Entered by: