Date: |
2016-02-09 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration was 09-17-2015
Cleaned cell before calibration
Cell T = 52.2, Pressure = 100.9
Zero Air offset = -20.0 railed
Span God CO2 421.22ppm offset = 11.28ppm
Zero 0.9477 -> 0.9423
Span 0.8394 -> 0.8220 |
Entered by: |
Joe Verfaillie |
|
|