Equipment and Calibration History

View Equipment |

View history entry 2126 for Licor LI-7500A sn: 75H-2150

Date: 2015-12-07
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2015-07-15 AGC=98.6 SDM=1 DAC1=- DAC2=- Ambient P=100.94kPa, Tair=19.78C, Tdew=12.44C Zero Air at 650psi, God CO2 421.22 at 700 Tdewspan=11.09 CO2 zero 0.9021 -> 0.9019 Offset: -1.16ppm CO2 span 1 -> 1.0114 Offset: -6.07ppm H2O zero 0.8671 -> 0.8671 Offset: 0.3ppm H2O span 0.9986 -> 1.0258 Offset: -0.52ppm Notes:
Entered by: