View history entry 2102 for Campbell Sci. CS511 sn: DO-8

Date: 2015-11-23
From: Lab - none
To: Lab - none
Title: Serviced
Notes: Serviced probe to go back in the field this week. The new calibration is: M = 9.45/36.1 in (mg/L)/mV (aka ppm/mV) (@ 18C).
Entered by: Sara Knox