Equipment and Calibration History

View Equipment |

View history entry 1950 for Licor LI-7700 sn: TG1-0223

Date: 2015-09-30
From: West Pond - Gas Analyzer CH4
To: West Pond - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: West Pond Last calibration: 2015-03-24 RSSI=66.5%, Optical RH=25% Ambient P=101.3kPa, Tair=18.1C Zero Air at 1050psi, AMP CH4 1.9966 at 1500 CH4 Zero Offset: 0.013ppm CH4 Span Offset: -0.0066ppm Notes: light, sporadic rain - no wind
Entered by:
Field notes