Equipment and Calibration History

View Equipment |

View history entry 1785 for Licor LI-7500A sn: 0041

Date: 2015-04-20
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2014-12-22 AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.14kPa, Tair=20.41C, Tdew=14.44C Zero Air at 900psi, God CO2 421.22 at 1200 Tdewspan=15.96 CO2 zero 1.0008 -> 1.0002 Offset: -2.0ppm CO2 span 1.0116 -> 1.0115 Offset: -0.02ppm H2O zero 0.8242 -> 0.8243 Offset: 0.8ppm H2O span 1.1234 -> 0.9991 Offset: 2.62ppm Notes: External serial port not working. Relatively large offset in H2O span, similar to other 7500s calibrated at this time - dew point generator and Vaisala probe agreed with in 0.5C
Entered by: